| VisuaLCN
Lab Client - Version Release Notes |
| |
Software Release 4.14b
Software Enhancements
Component |
Description |
|
Zero calibration reporting has been added for current and voltage. |
| Modules may now be calibrated with no zero voltage or current calibration points. |
Configure Utility |
Offline Circuit Recovery Period has been added to the Server Settings page. This allows a user to set the time in seconds that the server will wait between retrying to connect to offline circuits. |
Program Editor |
New safety feature: When saving a program, the software checks to make certain every step has a limit condition that will exit the step. If the step does not, a warning is displayed and the user can choose to save the program as is, or to cancel the save activity. |
| New safety feature: Two options can now be selected for current and power. When this feature is turned on, control parameters for these types will default to zero instead of the default specified in the module database. |
Code changes continually as features are added; unfortunately sometimes bugs are introduced and may not reside in earlier versions. It is recommended that you check the list that follows before requesting an update.
Bug Fixes
Component |
Description |
Server |
In Emergency Data Recording, the first data buffer dump would be lost. This problem has been fixed. |
Calibration 2 |
"Current As Found" was mistakenly divided by the scale factor. This has been fixed. |
| The current target as actually used is now saved into the data file and is editable. This value is used in the report, not the default value. |
Client |
The Program Map now correctly displays 2 more operators (equal and not equal) and 3 more limit actions (Pause Go To, Go To Subroutine, Return From Subroutine) when used with CAN variable limit conditions |
Program Editor |
The "Return From Subroutine" action for CAN limit conditions was fixed with the correct limit action code. |
Quick Data View |
The Quick Data View application will no longer crash when scrolling to a last page that is short enough not to require scroll bars. |
Data Export |
Columns for assignable data fields now export with the field names they've been assigned to, not "Assignable Variable 1", etc. This behavior matches that of the Quick Data View export. |
| Numbers behind decimal points were not recorded properly in all cases. This bug has been fixed. |
|
Software Release 4.13
Software Enhancements
Component |
Description |
Client |
VisuaLCN Lab Client now supports three new CAN control parameters: CAN Current, CAN Voltage and CAN power |
VisuaLCN Lab Client now supports subroutines |
Code changes continually as features are added; unfortunately sometimes bugs are introduced and may not reside in earlier versions. It is recommended that you check the list that follows before requesting an update.
Bug Fixes
Component |
Description |
Data Export |
Time scaling has been corrected |
Client |
Loss-of-Control has been added to the list of modes that allow cancellation of parallel circuit operation |
Program Editor |
When the CAN dictionary is changed, the program is marked as modified. The User will be prompted to save the program before it is closed. |
| The CAN variable reference counter will now increment correctly when a step or limit condition is copied. |
| Control Parameters and Limit Conditions with C-Rate and Number-of-Cell multipliers will now copy properly. They will also copy correctly during A-to-B module type conversion |
Quick Data View |
Headers for Assignable Data Channels will now sort in the correct order. Sorting is based on the new DataFieldNumber field in the test database. |
|
Software
Release 4.09b
Software Enhancements
| Component |
Description |
| Server |
VisuaLCN's
Server application new uses new syntax when creating default test
names. Instead of using a date/time stamp format, followed by a
circuit ordinal, Server new creates file names as date/time stamp,
followed by the circuit name. |
|
Software
Release 4.09
Software Enhancements
| Component |
Description |
| Program
Editor,
Client |
VisuaLCN's
Program Editor and Client applications now support C-rate programming. |
|
Software
Release 4.08
Software Enhancements
| Component |
Description |
| Program
Editor |
A
condensed spreadsheet format for printing program listing has been
added to the Program Editor application. |
|
Software
Release 4.07a
Code
changes continually as features are added; unfortunately sometimes bugs
are introduced and may not reside in earlier versions. It is recommended
that you check the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Server |
The
VisuaLCN Server no longer 'hangs' on shutdown under Windows 2000
or XP. This bug appeared in VisuaLCN versions 4.06a and 4.07. |
|
Software
Release 4.07
Software Enhancements
Code changes
continually as features are added; unfortunately sometimes bugs are introduced
and may not reside in earlier versions. It is recommended that you check
the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Quick
Data View |
Floating
point numbers longer than five digits will now display correctly. |
|
Software Release 4.06a
Software Enhancements
| Component |
Description |
| Client |
Lab
Client now supports up to 16 assignable data channels in support
of applicable Bitrode test circuits |
| Installation |
VisuaLCN
version upgrade installation is now fully automatic. Previous
version upgrades required the user to save settings and files
before manually removing oder versions, then restoring preference
settings and files. Version 4.06a upgrade installation, in most
cases, requires only that the user run the installation program. |
OS
Support |
With
the upgrade of the Windows Sockets version from 1.1 to 2.0, VisuaLCN
software, revision 4.06a or newer, will no longer run in a Windows
95 operating environment. Support for VisuaLCN products on the Windows
95 platform will continue until 15 February 2005. For more details
regarding system support, see Bitrode's
Software Support Policy. |
|
Software
Release 4.06
Software Enhancements
| Component |
Description |
| Client |
The
"Open Profile Database" dialog, activated either in the
Network Overivew or the Start Formation windows, now defaults to
the Most Recently Used file. |
| Quick
Data View |
An
alternative "one sheet wide" printing method has been
added, available as a choice in the Quick Data View print dialog.
Circuit data printed this way will appear in contiguous rows instead
of spanning pages horizontally. |
| Users
may choose either to print to a printer or to a file. This option
is available in the Quick Data View Print dialog. |
Quick
Graph |
The
text size of the graph title may now be changed in the Set Graph
Title window, selected from the menu bar or the toolbar. |
Code changes
continually as features are added; unfortunately sometimes bugs are introduced
and may not reside in earlier versions. It is recommended that you check
the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Client |
Users
may now resume from a fault condition on coldfire-based modules |
|
Software Release 4.05
Software Enhancements
| Component |
Description |
| Program
Editor |
VisuaLCN
Lab Client Program Editor has been updated to support global limit
conditions in the Program Headers. |
|
Software Release 4.04e
Code
changes continually as features are added; unfortunately sometimes bugs
are introduced and may not reside in earlier versions. It is recommended
that you check the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Data
Export, Quick Graph & Quick Data View |
These
applications scaled time to milliseconds when the time scale factor
was 0.01 for modules with 0.01 resolution, while the VisuaLCN Client
application was scaling the same time to seconds. This has been
corrected; Data Export, Quick Graph and Quick Data View now show
seconds for modules with 0.01 resolution. |
|
| |
Software
Release 4.05a
Code
changes continually as features are added; unfortunately sometimes bugs
are introduced and may not reside in earlier versions. It is recommended
that you check the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Quick
Data View |
The
"Apply an SQL Filter" toolbar button had different behavior
than the "Apply an SQL Filter" command in the pull down
menu or command button. Behaviors for these selections haves been
made consistent across the entire interface, giving the user the
choice of New, Edit or Apply from the Open/Save SQL Template window. |
Program
Editor |
Program
Headers can now be copied between test programs. |
|
Software Release 4.05
Software Enhancements
| Component |
Description |
| Program
Editor |
VisuaLCN
Lab Client Program Editor has been updated to support global limit
conditions in the Program Headers. |
|
Software Release 4.04e
Code
changes continually as features are added; unfortunately sometimes bugs
are introduced and may not reside in earlier versions. It is recommended
that you check the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Data
Export, Quick Graph & Quick Data View |
These
applications scaled time to milliseconds when the time scale factor
was 0.01 for modules with 0.01 resolution, while the VisuaLCN Client
application was scaling the same time to seconds. This has been
corrected; Data Export, Quick Graph and Quick Data View now show
seconds for modules with 0.01 resolution. |
|
Software Release 4.04b
Software Enhancements
| Component |
Description |
| Client |
User-defined
circuit names are now stored in the test database along with the
test data. |
| Quick
Data View |
A
"Circuit Name" field has been added to the Test Properties
dialog box, and is included in exported and printed data. |
| Data
Export |
Circuit
names are now included in exported data. |
|
Software Release 4.04a
Software Enhancements
| Component |
Description |
| User
Manual |
The
VisuaLCN User Manual is now in portable document format (PDF), included
on the VisuaLCN software installation disc. A self-extracting copy
of Adobe Reader 5.1 has been included on the VisuaLCN disc, and
is offered free of charge. |
| Quick
Data View |
Units
are now exported together with column headers |
| Test
information may now be printed before the test data, and test notes
may be printed after the test data. |
| When
tests are appended, an entry is added to the non-editable test notes,
indicating time, source database and the source test which was appended. |
| |
VisuaLCN
now supports assignable data channels and parallel operation of
multiple circuits. |
Code changes
continually as features are added; unfortunately sometimes bugs are introduced
and may not reside in earlier versions. It is recommended that you check
the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Program
Editor |
Steps
are now appended in numeric order for multiple-step cut and paste
operations, no matter the order in which they were selected. |
|
Software Release 4.03
Software Enhancements
| Component |
Description |
| Client |
A
new choice has been added to the Database Rollover Property Page
in the Configuration, allowing the display of a notification when
a test database reaches a specified size. |
| Quick
Data View |
Users
could choose a preferred time format setting (Seconds vs. HH:MM:SS)
when viewing or printing test data. Exported data now also honors
the time format preference. |
Code changes
continually as features are added; unfortunately sometimes bugs are introduced
and may not reside in earlier versions. It is recommended that you check
the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Program
Editor |
If
multiple steps were selected in the Program Steps window, a right-click
on a step would deselect all but one step. That has now been fixed;
all selected steps remain highlighted. |
|
Software Release 4.02
Code
changes continually as features are added; unfortunately sometimes bugs
are introduced and may not reside in earlier versions. It is recommended
that you check the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Lab
Client & Calibration |
Temperature
standby and chamber temperature modes are now supported.. |
| Calibration |
The
first time LCN Server was started after installation, a message
box appeared with the message "A sub key was unable to be
opened in the registry". If a user clicked the OK button,
the software ran properly, without any message box, from that
time forward. This has been modified; no message box appears,
though an entry will be made in the error log file. |
|
Software Release 4.01
Software
Enhancements
| Component |
Description |
| Quick
Data View |
Notes
and Editable Notes may now be exported with all other data to
the comma-delimited file. |
Code changes
continually as features are added; unfortunately sometimes bugs are introduced
and may not reside in earlier versions. It is recommended that you check
the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Client |
Module
type match check could only be disabled when a new test was started.
It can now also be disabled when a program is reloaded. |
| Data
Export |
Large
data sets would not successfully export. This has been corrected. |
|
Software Release 4.00
Software
Enhancements
| Component |
Description |
| Program
Editor |
The
user now has the option of selecting multiple program steps for
operations such as copying, cutting, pasting or deleting. |
| Data
Export |
Data
conversion may now be scheduled for automatic export at user-defined
periods. |
| Client |
Help
files have been updated with the mode icon description. Mode icons
are supported by tool tip notes. |
| Server |
NOTE:
For FCM modules only - User interface speed performance enhanced.
Users can now change the background of the circuit view on the
Loss of Control alarm. |
|
Software Release 3.39
Code
changes continually as features are added; unfortunately sometimes bugs
are introduced and may not reside in earlier versions. It is recommended
that you check the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Quick
Data View |
When
a source test was appended to a destination test, the time offset
would be incorrect in the destination record. Append Test is in
the Utilities menu - this has been corrected. |
| Server |
The
Server-to-Client communication would randomly prevent Client updates
- this has been corrected. |
|
Software Release 3.38a
Software Enhancements
| Component |
Description |
| Calibration |
Calibration
has been added as one of the components of VisuaLCN Lab Client. |
| Documentation |
In
VisuaLCN Lab/Formation Client a Documentation directory now exits
under the VisuaLCN directory structure, this is where the manual
resides. |
Code changes
continually as features are added; unfortunately sometimes bugs are introduced
and may not reside in earlier versions. It is recommended that you check
the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| l |
VisuaLCN
running under Windows 95/98/Me now properly displays bitmap images
which were missing in previous versions. |
Program
Editor |
The
scroll bar in the tree control of the Program Steps window was on
the left side of the window in Windows 2000/XP. The scroll bar is
now on the right side of the window, which is consistent with all
the other VisuaLCN components |
|
Software Release 3.37a
Code
changes continually as features are added; unfortunately sometimes bugs
are introduced and may not reside in earlier versions. It is recommended
that you check the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Quick
Data View |
The
application would crash if a read-only test database was opened.
Now Quick Data View can open a test database that has been configured
as a read only file. |
|
Software Release 3.36
Software Enhancements
| Component |
Description |
| Program
Editor |
A
new feature has been made available called Cell Volt Range. With
this feature a user can regulate the following limit conditions:
voltage, temperature, positive delta temperature or negative delta
temperature on the sequential selection of cell voltages. This is
an option that has to be added on a per module type basis. |
Code changes
continually as features are added; unfortunately sometimes bugs are introduced
and may not reside in earlier versions. It is recommended that you check
the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Client |
Test
Notes were not being added to database when initiated from Start
Test window, also various related code changes were incorporated
to the Test Notes area. |
| The
Program Map has been updated to more accurately format time values
as hours : minutes : seconds and to be consistent with Program Editor.
Previously, time values would show as hours : minutes : seconds
in Program Editor but would sometimes show up as seconds in Program
Map for the same module type. |
| If
the user was working with a module that had Pulse Time On and Pulse
Time Off variables, an error message would appear when the user
opened Program Map. These variables would not display correctly
in Program Map. |
| Quick
Data View |
By
default when exporting a test from the Test window the file name
was created using the test name. If there was a space in the test
name when the .csv file was created, the program creates the file
name only using the text up to the first space in the name and omitted
the remainder of the text causing Excel unable to open the file.
Exporting now allows spaces in test names. |
|
Software
Release 3.35a
Software Enhancements
| Component |
Description |
| Client |
Program
Map option has been modified to allows the user to see the program
steps of a program that has been downloaded to the module, if the
module has less than one hundred steps. This function displays all
the control parameters and limit conditions in a similair format
as viewed in Program Editor. The Program Map highlights the current
step of the running program. |
| Quick
Data View |
In
the Test Selection window, a new column has been added called Test
Date. Both Test Name and Test Date columns can be sorted ascending
or descending, the order is indicated by displaying a triangle in
either of those columns. |
| The
Test Date fields of the Test Properties and Test Selections windows
now show seconds in the date and time format. |
| If
a test to be Deleted or Moved has not run to completion, a message
box will show user a warning about possible data loss. |
Code changes
continually as features are added; unfortunately sometimes bugs are introduced
and may not reside in earlier versions. It is recommended that you check
the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Client |
In
the Test Notes window, if too much text was entered in the Notes
edit box, then text would scroll off the screen, the user was not
able to view all that was entered because there was no horizontal
scrollbar. |
| In
the Test Notes window, there was no limit to the amount of characters
typed into the text edit box. |
| The
"_" character can now be used when naming tests. |
| In
the Network Overview window Configuration tree if the root group
contained a child group and a child circuit, deletion of the child
group made the circuit a sibling of the root group when it should
have been unchanged (Lab Client only). |
| Server |
A
program could run even if one or more steps were rejected by the
module. This bug was introduced in version 3.34. |
| Program
Editor |
In
the Select Programs to Delete window, it is now possible to delete
individual programs or an empty group using the Delete button or
the Delete key on the keyboard. |
| Quick
Graph |
The
time display format in the User Settings dialog was initialized
to hours : minutes : seconds; however, the graph actually displayed
a time variable in seconds. This only happened initially until OK
or Apply has been clicked. |
| Quick
Data View |
The
Advanced Settings now retain the sort order when applied. |
|
Software
Release 3.34
Code changes continually as features are added; unfortunately sometimes
bugs are introduced and may not reside in earlier versions. It is recommended
that you check the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Client |
In
certain circumstances a mode was not updated when a module was started
or stopped. |
| Quick
Data View |
By
clicking the Extended Info and Clear All buttons and using the Advanced
Settings filters in certain sequences in the Test Selection dialog
window, an error message would display and the program would crash.
|
| The
Circuit Name drop-down list box in the Advanced Settings section
of the Test Selection dialog box would not show the correct information.
It's title has also been changed to Circuit Address. |
|
Software Release 3.33a
Software Enhancements
| Component |
Description |
| Client |
In
the Configure component, the "Data Acquisition Always Required"
Security is a new option. When set to True in the Security section,
the "Data Acquisition Required" checkbox in the Start dialog box
in Client is grayed out. The user is then not allowed to disable
data acquisition (Lab Client only). |
| Program
name and Circuit / Group name has been put on the Start Dialog title
bar. |
| Client
Control security was removed from the "Retreive Data From a Module"
operation. When Client Control security is set to TRUE it prohibits
another user from starting, stopping or reloading another program
in Client onto one started by the original user. However, it allows
the other user to see the data in Quick Data View or retreive data.
When disabled, it allows anyone to start, stop, resume the test,
reload another test or see the data in Quick Data View. Refer to
section 10 of the manual (Lab Client only). |
| Quick
Data View |
Export
Test Information, has been added to allows the user a choice of
exporting test information to a spreadsheet file along with the
data samples. |
Code changes
continually as features are added; unfortunately sometimes bugs are introduced
and may not reside in earlier versions. It is recommended that you check
the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Quick
Data View |
In
the Configure window, a user can select the Maximum Rows Per Export
File when exporting the data to a spreadsheet. The wrong number
of data samples, however, was being returned in the spreadsheet
files. |
| When
printing a single row, only one page would print when the row is
more than a page wide. This now prints as many pages as needed.
|
| The
Program section of the Test Properties window did not show the Program
Database when a test was selected. |
| The
Exclude feature was not functioning correctly. The number of records
chosen to be excluded from Quick Graph would often not be the ones
excluded. Instead, records from another test would be affected.
This has now been fixed so that the correct records chosen are excluded.
|
| Quick
Graph |
In
the Axis Properties window, the Grid Line Width list box which is
in the Major/Minor Grid Line groups, did not function correctly.
When a Grid Line Style other than solid was selected, the list box
is supposed to be disabled. However, it continued to be enabled
even after selecting a line other than solid. This behavior has
been corrected. |
| The
Axis Properties window did not populate the data showing the Maximum,
Minimum, Major Tick Space and Minor Ticks. After the OK button in
the Select Variables for Graph window is clicked and a new graph
is displayed, the Axis Properties window show the correct values.
|
| In
the New Graph window, where the graph is displayed, the X-axis started
at time of the first sample for "Total Time". It now starts at 0.
|
|
Software
Release 3.32
Software Enhancements
| Component |
Description |
| Client |
The
Selected Variables MRU feature was made optional. In the Select
Variables window for Individual and Group Status Views the user
can choose to keep the selection as a default, the next time Client
is started. |
Code changes
continually as features are added; unfortunately sometimes bugs are introduced
and may not reside in earlier versions. It is recommended that you check
the list that follows before requesting an update.
Bug Fixes
| Component |
Description |
| Client |
The
"Enter" key in the Start Test Window was a default key that would
start a new test if pressed, but has been disabled to prevent user
from mistakenly pressing that key which was the original design
(Lab Client only). |
| Program
Editor |
Program
Editor would not allow the creation of a new program database in
an empty root directory e.g. a:\. |
|
| |
| |
| |
| |