When running VisuaLCN, we cannot communicate with our Bitrode test
circuits.
Be certain that all of the circuit have the power turned on. Check
all BNC connectors to make sure they are tight and not dirty. Also
check the network adapter to see if both lights for transmit and receive
are blinking. If both are on continuously, check the corresponding
lights on each circuit to see if these circuit lights are also on
continuously. If so, cycle the power on the circuit and see if this
resolves the problem.
top |
After upgrading from LCN DOS to VisuaLCN, as well as the Network Adapter
between our LCN and the VisuaLCN computer, the Server component will
not communicate with the Network Adapter. Where should I look for
the problem?
Check the data rate on the Network Adapter. It may be set to a 9600
baud rate, too slow for the new network. 19.2K should be sufficient.
This can be found in the Configuration component of the VisuaLCN software.